Measurement of Spatial Stability in Precision Grip
Published on: June 4, 2020
Modernising grip dynamometry: Inter-instrument reliability between GripAble and Jamar
Sharah Abdul Mutalib1,2, Michael Mace3, Chloe Seager3
1GripAble Limited, London, UK. sharahamo@imperial.ac.uk.
The new GripAble device shows excellent reliability for measuring maximum grip strength (MGS). However, its readings are not interchangeable with the Jamar PLUS+ for patient monitoring.
Area of Science:
- Biomedical Engineering
- Rehabilitation Technology
- Clinical Measurement
Background:
- Maximum grip strength (MGS) is a key indicator of overall health and physiological status.
- Accurate and reliable MGS measurement tools are essential for valid health assessments.
- The GripAble, a mobile hand grip device, has been developed for MGS assessment.
Purpose of the Study:
- To evaluate the performance of the GripAble device for measuring maximum grip strength.
- To assess the inter-instrument reliability of GripAble against the established Jamar PLUS+ dynamometer.
Main Methods:
- Sixty-three participants (N=63) provided MGS data from both hands using GripAble and Jamar PLUS+.
- Inter-instrument reliability was analyzed using Intraclass Correlation (ICC), regression, and Bland-Altman analysis.
Main Results:
- GripAble demonstrated good-to-excellent inter-instrument reliability with the Jamar PLUS+ (ICC3,1 = 0.906).
- GripAble measurements were equivalent to 69% of Jamar PLUS+ measurements.
- A proportional difference was observed, with MGS differences increasing with higher grip strength readings.
Conclusions:
- GripAble is a reliable tool for measuring grip strength.
- Direct interchange or translation of MGS readings between GripAble and Jamar PLUS+ is not recommended for serial measurements.
- Future work includes collecting normative MGS data for GripAble to facilitate comparisons.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:35Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor
Published on: June 16, 2021
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Electronic Distance Measuring Instruments
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Uncertainty in Measurement: Accuracy and Precision
Calibration Curves: Correlation Coefficient