Modernising grip dynamometry: Inter-instrument reliability between GripAble and Jamar
Sharah Abdul Mutalib1,2, Michael Mace3, Chloe Seager3
1GripAble Limited, London, UK. sharahamo@imperial.ac.uk.
Introduction:
Maximum grip strength (MGS) is a reliable biomarker of overall health and physiological well-being. Therefore, an accurate and reliable measurement device is vital for ensuring the validity of the MGS assessment. This paper presents GripAble, a mobile hand grip device for the assessment of MGS. GripAble's performance was evaluated using an inter-instrument reliability test against the widely used Jamar PLUS+ dynamometer.
Methods:
MGS data from sixty-three participants (N = 63, median (IQR) age = 29.0 (29.5) years, 33 M/30 F) from both hands using GripAble and Jamar PLUS+ were collected and compared. Intraclass correlation (ICC), regression, and Bland and Altman analysis were performed to evaluate the inter-instrument reliability and relationship in MGS measurements between GripAble and Jamar PLUS+ .
Results:
GripAble demonstrates good-to-excellent inter-instrument reliability to the Jamar PLUS+ with ICC3,1 = 0.906 (95% CI [0.87-0.94]). GripAble's MGS measurement is equivalent to 69% (95% CI [0.67-0.71]%) of Jamar PLUS+'s measurement. There is a proportional difference in mean MGS between the two devices, with the difference in MGS between GripAble and Jamar PLUS+ increasing with MGS.
Conclusion:
The GripAble is a reliable tool for measuring grip strength. However, the MGS readings from GripAble and Jamar PLUS+ should not be interchanged for serial measurements of the same patient, nor be translated directly from one device to the other. A new normative MGS data using GripAble will be collected and accessed through the software for immediate comparison to age and gender-matched subpopulations.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:35Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor
Published on: June 16, 2021
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Electronic Distance Measuring Instruments
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Uncertainty in Measurement: Accuracy and Precision
Calibration Curves: Correlation Coefficient