Can a humid storage environment of surgical instruments before reprocessing increase patient safety and durability of
P Rubak1, J Lorenzen2, K Ripadal3
1Clinic for Diagnostics & Clinical Nursing Research Unit, Aalborg University Hospital, Aalborg, Denmark.
Background:
National and international guidelines recommend reprocessing of medical instruments to commence as soon as possible post-surgery; furthermore, they recommend that transport and storage of surgical instruments postoperatively occurs in a moist, humid atmosphere. The concern is that a dry storage environment results in deterioration of instruments.
Aim:
To evaluate whether residual protein or corrosion is associated with storage environment (dry or humid), holding time or number of treatment cycles.
Methods:
The range of protein residue and corrosion were tested on surgical instruments contaminated with human blood amended Enterococcus faecalis ATCC 29212. Subsequently instruments were stored for 6, 12 and 24 h in dry or humid conditions. After one, 25 and 50 reprocessing cycles, instruments were examined for protein residues using the o-phthaldialdehyde (OPA) method or corrosion using stereomicroscopy, scanning electron microscopy and energy dispersive spectroscopy.
Findings:
Protein residue found on instruments was 21.5-54.0 μg and corrosion corresponded to 0-5% of the inspected area. No associations between storage environment and protein residue (adjusted mean difference = 0.48, 95% confidence interval: -0.42, 1.37, P=0.30) or corrosion (P=0.20) were identified. Higher numbers of treatment cycles showed higher amounts of corrosion (mean: 1cycle = 0.06%, 25cycles = 0.52% and 50cycles = 1.45%). In contrast, higher numbers of treatment cycles showed lower amounts of protein residue (P<0.001). We found both lower protein residue concentration and lower corrosion rating at 12 h compared with 6 and 24 h holding time.
Conclusion:
Cleanliness and durability of instruments before reprocessing seems not to be affected by storage environment or holding time but instead by number of treatment cycles.
More Related Videos
10:52Preparation and High-temperature Anti-adhesion Behavior of a Slippery Surface on Stainless Steel
Published on: March 29, 2018
07:15A Testing Platform for Durability Studies of Polymers and Fiber-reinforced Polymer Composites under Concurrent Hygrothermo-mechanical Stimuli
Published on: December 11, 2014
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Key Techniques in Microbiology
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...
Asepsis