Verification of Assayed Blood Gas Quality Control Ranges.
Jeffrey M Haynes1, Robert G Fishwick2
1Pulmonary Function Laboratory, Elliot Health System, Manchester, New Hampshire. jhaynes@elliot-hs.org.
Laboratory quality control (QC) for blood gas analysis requires verifying manufacturer ranges. This study found manufacturer ranges were often too wide and unacceptable for clinical use, emphasizing the need for laboratory verification.
Area of Science:
- Clinical Laboratory Science
- Medical Diagnostics
- Analytical Chemistry
Background:
- Blood gas quality control (QC) is a mandatory laboratory process.
- Laboratories must verify manufacturer-provided QC ranges, a Clinical Laboratory Improvement Amendment (CLIA) requirement.
- Acceptable QC ranges are typically 2 standard deviations (SD) from the mean.
Purpose of the Study:
- To verify the mean and 2 SD ranges of QC measurements from a blood gas analyzer.
- To assess the agreement between laboratory-measured QC data and manufacturer-provided data.
- To evaluate the clinical acceptability of manufacturer-provided QC ranges.
Main Methods:
- Compared QC data from a blood gas analyzer to manufacturer-provided mean and ranges.
- Calculated the percent difference between measured and manufacturer-provided mean values.
- Determined how many SD the manufacturer ranges were from the measured mean using measured SD.
Main Results:
- The largest mean value difference was 2.27% above the manufacturer mean.
- 48% of mean comparisons showed 0% difference, and 71% were less than 1%.
- Manufacturer ranges (2.4-75 SD) were significantly wider than measured 2 SD ranges and deemed unacceptable for clinical use.
Conclusions:
- Analysis validates the CLIA mandate for QC range verification.
- Laboratories must verify and adjust QC means and ranges to match their specific blood gas analyzer's performance.
- This ensures accurate and reliable blood gas testing in clinical settings.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Quality Assurance
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include: