Verification of Assayed Blood Gas Quality Control Ranges
Jeffrey M Haynes1, Robert G Fishwick2
1Pulmonary Function Laboratory, Elliot Health System, Manchester, New Hampshire. jhaynes@elliot-hs.org.
Background:
Blood gas quality control (QC) is an essential and mandatory part of a laboratory's quality plan. The acceptable QC range should be 2 SD from the mean value. The use of assayed QC material does not negate the responsibility of the laboratory to calculate the mean and 2 SD ranges of QC measurements for verification. Verifying assayed QC ranges is a Clinical Laboratory Improvement Amendment (CLIA) requirement. This study shows the results of assayed QC mean and 2 SD range verification from a blood gas analyzer.
Methods:
QC data from a blood gas analyzer were compared to manufacturer-provided mean and ranges. The percent difference between the measured mean and the manufacturer-provided mean was calculated to assess agreement. The measured SD was used to determine how many SD the manufacturer-provided ranges were from the measured mean.
Results:
The largest difference in mean values was 2.27% > the manufacturer-provided mean. Forty-eight percent of all mean value comparisons showed a difference of 0%, and 71% were < 1%. The manufacturer-provided ranges were considerably wider than the measured 2 SD range, ranging from 2.4-75 SD. None of the manufacturer-provided ranges were deemed acceptable for clinical use.
Conclusions:
Our analysis validates the CLIA mandate and American Association for Respiratory Care Clinical Practice Guideline recommendation that laboratories must verify manufacturer-provided QC means and ranges and adjust QC means and ranges to match the performance of their blood gas analyzer.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Quality Assurance
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include: