Related Experiment Video
Updated: Oct 5, 2025

Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Comparing Central Aortic Pressures Obtained Using a SphygmoCor Device to Pressures Obtained Using a Pressure Catheter
Cara Esposito1, Priscilla Machado1, Ira S Cohen2
1Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
This study shows that using a linear-fit function significantly improves the accuracy of noninvasive aortic pressure estimations from the SphygmoCor device, making them comparable to invasive catheter measurements.
Area of Science:
- Cardiovascular Physiology
- Medical Device Technology
- Diagnostic Accuracy
Background:
- Noninvasive estimation of central aortic pressure is crucial for cardiovascular assessment.
- The SphygmoCor XCEL Pulse Wave Analysis (PWA) device provides noninvasive pressure estimations.
- Validation against invasive catheterization is necessary to establish clinical utility.
Purpose of the Study:
- To compare noninvasive aortic pressures estimated by the SphygmoCor XCEL device with invasive measurements.
- To evaluate the impact of a linear-fit function on improving the accuracy of these estimations.
- To determine if SphygmoCor estimations can be made comparable to gold-standard catheter measurements.
Main Methods:
- 136 patients undergoing cardiac catheterization were enrolled.
- Aortic pressures were measured invasively via catheter and noninvasively using SphygmoCor.
- Bland-Altman analysis and a novel linear-fit function were applied to assess and correct errors.
Main Results:
- Initial errors between measured and estimated aortic pressures were substantial (mean errors >6.4 mm Hg).
- Application of the linear-fit function significantly reduced these errors to less than 1 mm Hg.
- Bootstrap analysis confirmed the generalizability of the linear-fit function.
Conclusions:
- The SphygmoCor XCEL device can provide accurate aortic pressure estimations when combined with the derived linear-fit function.
- This method enhances the comparability of noninvasive estimations to invasive pressure catheter measurements.
- The findings support the clinical use of SphygmoCor data with the linear-fit adjustment for improved diagnostic accuracy.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:

