COMPARISON OF OPTICAL BIOMETERS ARGOS AND IOL MASTER 700
Purpose:
To compare the biometric data obtained by the new optical biometer Argos and the conventionally used biometer IOL Master 700.
Patients And Methods:
Retrospective analysis of the biometric data of 57 patients (106 eyes) who were examined at TANA Ophthalmology Clinic s.r.o in Olomouc. Patient measurements were carried out on both devices on the same day by the same optometrist within the standard preoperative calculation of the intraocular lens before cataract surgery. Evaluated and statistically analyzed biometric data were axial lenght, anterior chamber depth, average keratometry and lens thickness.
Results:
The correlation between all compared data was high, with statistical significance p < 0.01. Bland-Altman plots showed good agreement with a 95% limit of agreement. Axial lenght, average keratometry and lens thickens did not show significant differences (p = 0.941; p = 0.773; p = 0.860). IOL Master 700 showed flatter average keratometry, however, the differences were numerically small and insignificant. Anterior chamber depths obtained by Argos were longer, with a significance p < 0.05.
Conclusion:
The segmental refractive index technology used by Argos caused differences in anterior chamber depths. Overall axial length was, however, not, in our cohort of patients, affected by this. In general, the optical biometers Argos and IOL Master 700 show excellent agreement in the measured biometric data.