The Perils of Buddy Wire Use With a Filterwire
Alessandra Gutierrez1, Yashasvi Chugh1, Spyridon Kostantinis1
1Minneapolis Heart Institute and Minneapolis Heart Foundation, Abbott Northwestern Hospital, Minneapolis, MN, USA.
Abstract:
Embolic protection devices may decrease the risk of distal embolization during percutaneous coronary intervention of saphenous vein graft and highly thrombotic lesions. Use of a buddy wire can facilitate equipment delivery but carries perils when used together with a filter. We present two cases demonstrating potential complications of buddy wire use with a filter. In the first case a saphenous vein graft perforation occurred after stenting. A buddy wire and second balloon were used to maintain hemostasis while removing the Filterwire. The buddy wire, however, was caught in the filter retrieval catheter, interlocking the two guide catheters and was eventually removed after forceful pulling. In the second case a Filterwire was used to prevent distal embolization of a highly thrombotic right coronary artery lesion. A buddy wire was used to assist with equipment delivery. A stent was inadvertedly deployed over the buddy wire, entrapping the Filterwire. The entrapped Filterwire was eventually removed after multiple balloon inflations over the buddy wire.
Related Concept Videos
Magnetic Force On Current-Carrying Wires: Example
Magnetic Field Due to Two Straight Wires
Line Protection with Impedance Relays
Under normal conditions, low load currents keep the measured...
Cable Subjected to Concentrated Loads
Magnetic Field Due To A Thin Straight Wire
Circuit Breaker and Fuse Selection
In high-voltage systems,...