Technical note: The impact of storage humidity on the response of reference-class ionization chambers
Yuichiro Morishita1, Morihito Shimizu1, Nobuhiro Takase2
1National Metrology Institute of Japan, AIST, Tsukuba, Ibaraki, Japan.
Purpose:
A number of Farmer-type ionization chambers were tested against storage humidity to confirm whether they satisfied the criteria for the long-term stability of reference-class ionization chambers.
Methods:
The ionization chambers were stored for several months in an environment with relative humidity adjustable from 20% to 80%. The ionization chambers were removed from the storage environment at variable intervals ranging from 1 to 70 days and irradiated in a Co-60 radiation beam. The responses for each ionization chamber were evaluated from the measured currents corrected for the atmospheric air density, and were compared with those predicted by the Co-60 half-life.
Results:
Certain ionization chambers gave a constant relative response regardless of the storage humidity, while the relative responses of two types of ionization chambers changed as a function of the storage humidity. The difference between the relative responses for the low (20-30%) and high (70-80%) storage relative humidity was 0.7%. The response was larger for the high relative humidity storage. Immediately after the storage humidity changed, the relative response started to change by the day, and it took approximately 2 weeks to 2 months for the relative response to converge. For one type of the ionization chamber, the plastic outer wall and the outer electrode were replaced with those made of solid graphite, and it was confirmed that the remodeled ionization chamber did not exhibit the response change.
Conclusions:
The present results and previous reports by other authors indicate that the magnitude of the change depends on the magnitude of the water absorption of the plastic used for the outer wall and/or the electrode of the ionization chamber. Thus, it is important in the selection of the reference-class ionization chamber to note the material and structure of the outer wall and electrode of the ionization chamber. If the ionization chamber has a hygroscopic wall and electrode and it is used as a reference ionization chamber, it is necessary to pay additional attention to the humidity difference for the storage, daily irradiation, and yearly calibration especially in regions with large seasonal humidity fluctuations.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
07:44Determining the Chemical Composition of Corrosion Inhibitor/Metal Interfaces with XPS: Minimizing Post Immersion Oxidation
Published on: March 15, 2017
Related Concept Videos
Chemical Ionization (CI) Mass Spectrometry
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...