Improving image quality by optimizing beam width and helical pitch in CT
Related Experiment Video
Updated: Oct 5, 2025

Three-Dimensional Imaging of Tumor-Bearing Tissue Using the Iterative Bleaching Extends Multiplexity Approach
Published on: April 25, 2025
Pil-Hyun Jeon1, Chang-Lae Lee2
1Department of Diagnostic Radiology, Yonsei University Wonju College of Medicine, Wonju Severance Christian Hospital, Wonju-si, Gangwon-do, Republic of Korea.
Background:
Expanding computed tomography (CT) detector coverage broadens the beam width, but inaccurate tube current application can reduce image quality at the boundaries between body regions with different attenuation values along the z-axis.
Objective:
This study aims to develop and validate a new CT scanning technique with a fixed pitch to achieve higher imaging quality.
Methods:
A cylindrical water phantom and an anthropomorphic chest phantom with different diameters represent a human body with different attenuation values. By optimizing the beam width and helical pitch, the pitch is fixed during scanning. The mean noise of the images and the standard deviation were calculated, and the coefficient of variation (COV) was compared to evaluate the uniformity of image noise according to the beam width.
Results:
At the boundaries between regions with different attenuation values, the 10 mm beam width (COV: 0.065) in the water phantom showed a 47.7% COV reduction of image noise compared with the 20 mm beam width (COV: 0.125). In addition, the 20 mm beam width (COV: 0.146) in the chest phantom showed a 29.3% COV reduction of image noise compared with the 40 mm beam width (COV: 0.206). Thus, as the beam was narrowed, the mean noise was similar, but the standard deviation was reduced.
Conclusions:
The proposed CT scanning technique with a fixed pitch, optimized beam width, and helical pitch demonstrates that image quality can be improved without increasing radiation dose at the boundary between regions with different attenuation values.
More Related Videos
11:49Using High Resolution Computed Tomography to Visualize the Three Dimensional Structure and Function of Plant Vasculature
Published on: April 5, 2013
08:46Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
Published on: April 13, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...