Published on: December 3, 2016
A systematic study on effects of calibration-waveguide geometry and least-squares formulation on ear-probe source
Kren Monrad Nørgaard1, Joshua J Hajicek2
1Interacoustics A/S, Audiometer Allé 1, Middelfart, DK-5500, Denmark.
Abstract:
Measuring ear-canal absorbance and compensating for effects of the ear-canal acoustics on otoacoustic-emission measurements using an ear probe rely on accurately determining its acoustic source parameters. Using pressure measurements made in several rigid waveguides and models of their input impedances, a conventional calibration method estimates the ear-probe Thévenin-equivalent source parameters via a least-squares fit to an over-determined system of equations. Such a calibration procedure involves critical considerations on the geometry and number of utilized calibration waveguides. This paper studies the effects of calibration-waveguide geometry on achieving accurate ear-probe calibrations and measurements by systematically varying the lengths, length ratios, radii, and number of waveguides. For calibration-waveguide lengths in the range of 10-60 mm, accurate calibrations were generally obtained with absorbance measurement errors of approximately 0.02. Longer waveguides resulted in calibration errors, mainly due to coincident resonance frequencies among waveguides in the presence of mismatches between their assumed and actual geometries. The accuracy of calibrations was independent of the calibration-waveguide radius, except for an increased sensitivity of wider waveguides to noise. Finally, it is demonstrated how reformulating the over-determined system of equations to return the least-squares reflectance source parameters substantially reduces calibration and measurement errors.
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Correlation Coefficient
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Influence of Earth's Curvature and Atmospheric Refraction on Leveling