Percutaneous Ventricular Restoration Prevents Left Ventricular Remodeling Post Myocardial Infarction: One-Year

Zhengbin Zhu1, Jinzhou Zhu1, Jiwei Yu2

  • 1Department of Cardiology, Rui Jin Hospital, Shanghai Jiao Tong University School of Medicine. Shanghai, China; Cardiovascular Research institution, Shanghai Jiao Tong University School of Medicine. Shanghai, China.

Insights

The Heartech left ventricular partitioning device (LVPD) shows promise for treating chronic heart failure (HF) post-myocardial infarction. This novel therapy effectively reduced heart size and improved function in patients, demonstrating safety and efficacy up to 12 months.

Area of Science:

  • Cardiology
  • Medical Devices
  • Regenerative Medicine

Background:

  • Chronic heart failure (HF) following myocardial infarction (MI) remains a significant clinical challenge.
  • The Heartech left ventricular partitioning device (LVPD) is a novel implantable therapy designed for ventricular restoration.
  • This study represents the first human investigation of the Heartech LVPD.

Purpose of the Study:

  • To evaluate the safety and efficacy of the Heartech LVPD in patients with chronic HF post-MI.
  • To assess the impact of LVPD implantation on left ventricular (LV) volumes, LV ejection fraction (LVEF), and functional status.
  • To document major adverse cardiovascular and cerebrovascular events (MACCEs) following the procedure.

Main Methods:

  • A prospective, first-in-human study involving 16 patients from three cardiac centers in China.
  • Percutaneous implantation of the Heartech LVPD.
  • Comprehensive assessments including echocardiography, European Quality of Life-5 Dimensions (EQ-5D) scale, and 6-minute walk test (6MWT) at baseline and 12-month follow-up.

Main Results:

  • Successful LVPD implantation in 93.75% of patients.
  • Significant reduction in LV end-systolic volume index (LVESVI) and LV end-diastolic volume index (LVEDVI) (P=.001 and P=.002, respectively).
  • Significant improvement in LVEF (P=.003), 6MWT distance (P=.01), and EQ-5D scores (P<.001).
  • One patient experienced a fatal MI; no other MACCEs were reported at 12 months.

Conclusions:

  • The Heartech LVPD is safe and effective for reducing LV volume and improving LV function in patients with chronic HF post-MI.
  • The observed benefits in cardiac structure and function were sustained up to the 12-month follow-up.
  • The Heartech LVPD represents a promising therapeutic option for managing heart failure after myocardial infarction.