Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific
Anton Mans1, Roel Rozendaal1, Tomas Janssen1
1Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, The Netherlands.
Background And Purpose:
Dosimetric patient-Specific Quality Assurance (PSQA) data contain in addition to cases with alerts, many cases without alerts. The aim of this study was to present a procedure to investigate long-term trend analysis of the complete set of PSQA data for the presence of site-specific deviations to reduce underlying systematic dose uncertainties.
Materials And Methods:
The procedure started by analysing a large set of prostate Volumetric Modulated Arc Therapy (VMAT) PSQA data obtained by comparing 3D electronic portal image device (EPID)_based in vivo dosimetry measurements with dose values predicted by the Treatment Planning System (TPS). If systematic deviations were present, several actions were required. These included confirmation of these deviations with an independent dose verification system for which a 2D detector array in a phantom was used, and analysing calculated with measured PSQA data, or delivery machine characteristics. Further analysis revealed that the under-dosage correlated with plan complexity and coincided with changes in clinically applied planning techniques.
Results:
Prostate VMAT PSQA data showed an under-dosage gradual increasing to about 2% in 3 years, which was confirmed by the measurements with the 2D detector array in a phantom. The implementation of new beam fits in the TPS led to a reduction of the observed deviations.
Conclusion:
Long-term analysis of site-specific PSQA data is a useful method to monitor incremental changes in a radiotherapy department due to various changes in the treatment planning and delivery of prostate VMAT, and may lead to a reduction of systematic dose uncertainties in complex treatments.