Sheath shape pattern during leadless pacemaker implantation
Motomi Tachibana1, Kimikazu Banba2, Yusuke Hasui2
1Department of Cardiology, Sakakibara Heart Institute of Okayama, 2-5-1 Nakai-cho, Kita-ku, Okayama, 700-0804, Japan. motomitachibana@yahoo.co.jp.
Summary
Predicting leadless pacemaker (LP) sheath shape using CT scans can optimize implantation. Narrower angles between the inferior vena cava (IVC) and interventricular septum (IVST) and right ventricular apex (RVA) indicate a need for an alpha-loop sheath.
Area of Science:
- Cardiology
- Medical Imaging
- Biomedical Engineering
Background:
- Leadless pacemakers (LPs) offer a less invasive approach to bradyarrhythmia treatment.
- Current methods for shaping LP delivery sheaths lack patient-specific anatomical customization.
- Optimizing sheath shape is crucial for successful and safe LP implantation.
Purpose of the Study:
- To predict the required shape of the leadless pacemaker (LP) delivery sheath during implantation.
- To correlate preoperative computed tomography (CT) measurements with intraoperative sheath configuration.
- To identify anatomical predictors for the alpha-loop sheath shape.
Main Methods:
- Retrospective analysis of 48 patients who received LPs.
- Categorization into alpha-loop and non-alpha-loop groups based on sheath head shape.
- Measurement of angles between the inferior vena cava (IVC) and interventricular septum (IVST), and IVC and right ventricular apex (RVA) using CT.
- Assessment of sheath position relative to a vertical line in fluoroscopy (LAO view).
Main Results:
- Significantly smaller IVC-IVST and IVC-RVA angles were observed in the alpha-loop group (p < 0.01).
- A combined IVC-IVST angle < 50° and IVC-RVA angle < 55° predicted alpha-loop shape with 81.8% sensitivity.
- The delivery sheath was more frequently positioned to the right of the vertical line in the alpha-loop group (90.9% vs. 23.1%, p < 0.01).
Conclusions:
- Preoperative CT-derived angles (IVC-IVST and IVC-RVA) can predict the need for an alpha-loop sheath.
- Narrower preoperative angles suggest the necessity of shaping the delivery sheath head into an alpha-loop.
- This prediction aids in tailoring the delivery sheath to individual cardiac anatomy for LP implantation.