787
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
787
Tandem Mass Spectrometry01:21

Tandem Mass Spectrometry

1.4K
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and signal-to-noise ratio for the analyte. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
1.4K
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences01:17

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

965
A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
965

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Digital Revitalization of a Legacy Linear Ion Trap System.

Analytical chemistry·2026
Same author

Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry.

Journal of the American Society for Mass Spectrometry·2025
Same author

Phase Modulation to Increase Ion Throughput for Ion Mobility-Time-of-Flight Experiments.

Journal of the American Society for Mass Spectrometry·2025
Same author

A Hybrid Vacuum Flange RF Oscillator for Low-Cost Mass Spectrometry.

Journal of the American Society for Mass Spectrometry·2025
Same author

Efficient Coupling of Structures for Lossless Ion Manipulations with Ion Trap Mass Analyzers Using Phase Modulation.

Journal of the American Society for Mass Spectrometry·2025
Same author

Advancing Protein Analysis: A Low-Pressure Drift Tube Orbitrap Mass Spectrometer for Ultraviolet Photodissociation-Based Structural Characterization.

Analytical chemistry·2024

Related Experiment Video

Updated: Oct 4, 2025

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
08:40

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments

Published on: January 20, 2022

4.5K

Synchronized Stepped Frequency Modulation for Multiplexed Ion Mobility Measurements.

Elvin R Cabrera1, Brian H Clowers1

  • 1Department of Chemistry, Washington State University, Pullman, Washington 99164, United States.

Journal of the American Society for Mass Spectrometry
|February 2, 2022
PubMed
Summary
This summary is machine-generated.

A new discrete stepwise frequency modulation synchronizes ion mobility spectrometry with mass analyzers, eliminating data ambiguity. This method enhances signal-to-noise ratios for improved analyte detection in mass spectrometry.

Keywords:
Fourier transformion mobilityion trapmass spectrometrymultiplexing

More Related Videos

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
08:22

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization

Published on: August 6, 2018

7.0K
T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis
16:40

T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis

Published on: July 31, 2010

24.8K

Related Experiment Videos

Last Updated: Oct 4, 2025

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
08:40

Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments

Published on: January 20, 2022

4.5K
Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
08:22

Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization

Published on: August 6, 2018

7.0K
T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis
16:40

T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis

Published on: July 31, 2010

24.8K

Area of Science:

  • Analytical Chemistry
  • Spectrometry

Background:

  • Frequency-encoded multiplexing enables direct coupling of ion mobility spectrometry (IMS) with mass analyzers.
  • Traditional continuous frequency sweeps in Fourier transform IMS (FT-IMS) cause ambiguity due to duty cycle mismatch with ion-trap mass analyzers.

Purpose of the Study:

  • To eliminate ambiguity in FT-IMS by synchronizing frequency modulation with ion-trap mass analyzer scan rates.
  • To introduce a discrete stepwise frequency modulation method for improved data acquisition.

Main Methods:

  • Implemented a discrete stepwise function to modulate ion gates in IMS.
  • Achieved synchronization between generated frequencies and the linear ion trap scan rate.
  • Compared performance against continuous linear modulation techniques.

Main Results:

  • Eliminated ambiguity in assigning ion current to specific modulation frequencies.
  • Maintained high precision across various terminal frequencies and scan lengths.
  • Demonstrated no loss in performance compared to continuous modulation despite fewer frequencies.

Conclusions:

  • Discrete stepwise frequency modulation offers a precise and unambiguous method for FT-IMS coupled with ion-trap mass analyzers.
  • Frequency-scan synchronization enables advanced data processing, significantly improving signal-to-noise ratios.
  • This approach enhances detection capabilities for analytes of varying intensities.