Acute intraocular toxicity caused by perfluorocarbon liquids: safety control systems of medical devices
Girish K Srivastava1,2, Vivekanandan Kalaiselvan3, Cristina Andrés-Iglesias4
1Instituto Universitario de Oftalmobiología Aplicada (IOBA), Universidad de Valladolid, Valladolid, Spain. girish@ioba.med.uva.es.
Background:
Acute retinal toxicity has been demonstrated to be associated with the intraoperative use of perfluorocarbon liquids (PFCLs), especially perfluorooctane (PFO). Recently, several cases of PFO-associated blindness have been reported in Spain, Holland, France, Italy, the Middle East, and South America.
Methods:
As a result, a new ISO guideline (ISO 16672:2020) was drafted, discussed, approved, and released in 2019. This recent ISO16672:2020 guideline recommends performing direct cytotoxicity tests as an option along with chemical analysis to measure PFCL quality (purity and safety).
Results:
In this review paper, it has been emphasized why an appropriate biological test, specifically direct exposure of PFCL to live cells, for measuring cytotoxicity must be performed with each PFCL batch along with chemical analysis.
Conclusions:
The paper intends to compile all available information to discuss possible approaches for avoiding adverse clinical cases in future.
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
11:49Synthesis of Soft Polysiloxane-urea Elastomers for Intraocular Lens Application
Published on: March 8, 2019
Related Concept Videos
Glaucoma: Overview
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications