Simulation of thermal sanitization of air with heat recovery as applied to airborne pathogen deactivation
M Busto1, E E Tarifa2, M Cristaldi3
1Institute of Research On Catalysis and Petrochemistry, INCAPE, FIQ-UNL, CONICET, Collecting Ring, National Road 168 km 0, El Pozo, 3000 Santa Fe, Argentina.
Abstract:
The technique of air sterilization by thermal effect was revisited in this work. The impact of incorporating a high efficiency heat recovery exchanger to a sterilizing cell was especially assessed. A mathematical model was developed to study the dynamics and the steady state of the sterilizer. Computer simulation and reported data of thermal inactivation of pathogens permitted obtaining results for a proof-of-concept. The simulation results confirmed that the incorporation of a heat recovery exchanger permits saving more than 90% of the energy needed to heat the air to the temperature necessary for sterilization, i.e., sterilization without heat recovery consumes 10-20 times the energy of the same sterilization device with heat recovery. Sanitization temperature is the main process variable for sanitization, a fact related to the activated nature of the thermal inactivation of viruses and bacteria. Heat recovery efficiency was a strong function of the heat transfer parameters but also rather insensitive to the cell temperature. The heat transfer area determined the maximum capacity of the sterilizer (maximum air flowrate) given the restrictions of minimum sanitization efficiency and maximum power consumption. The proposed thermal sterilization device has important advantages of robustness and simplicity over other commercial sterilization devices, needing practically no maintenance and eliminating a big variety of microorganisms to any desired degree. For most pathogens, the inactivation can be total. This result is not affected by the uncertainties in thermal inactivation data, due to the Arrhenius-like dependence of inactivation. Temperature can be adjusted to achieve any removal degree.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Temperature
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...