Purpose:
The GammaPod is a novel device for stereotactic breast treatments that employs 25 rotating Co-60 sources while the patient is continuously translated in three axes to deliver a highly conformal dose to the target. There is no commercial software available for independent second calculations. The purpose of this study is to determine an efficient way to estimate GammaPod treatment times based on target volume and use it as a second calculation for patient-specific quality assurance.
Methods:
Fifty-nine GammaPod (Xcision Medical Systems, LLC.) breast cancer patient treatments were used as the fitting dataset for this study. Similar to the Curie-seconds concept in brachytherapy, we considered dose-rate × time/(prescribed dose) as a function of target volumes. Using a MATLAB (Mathworks, Natick, MA, USA) script, we generated linear (with 95% confidence interval (CI)) and quadratic fits and tested the resulting equations on an additional set of 30 patients.
Results:
We found a strong correlation between the dose-rate × time/(prescribed dose) and patients' target volumes for both the linear and quadratic models. The linear fit was selected for use and using the polyval function in MATLAB, a 95% CI graph was created to depict the accuracy of the prediction for treatment times. Testing the model on 30 additional patients with target volumes ranging from 20 to 188 cc yielded treatment times from 10 to 25 min that in all cases were within the predicted CI. The average absolute difference between the predicted and actual treatment times was 1.0 min (range 0-3.3 min). The average percent difference was 5.8% (range 0%-18.4%).
Conclusion:
This work has resulted in a viable independent calculation for GammaPod treatment times. This method has been implemented as a spreadsheet that is ready for clinical use to predict and verify the accuracy of breast cancer treatment times.
More Related Videos
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Drug Accumulation During Multiple Dosing: Repetitive IV Injections
Dose Size and Dosing Frequency: Determination Methods