Validity of continuous glucose monitoring for categorizing glycemic responses to diet: implications for use in
Jordi Merino1,2,3, Inbar Linenberg4, Kate M Bermingham5,6
1Diabetes Unit and Center for Genomic Medicine, Massachusetts General Hospital, Boston, MA, USA.
Background:
Continuous glucose monitor (CGM) devices enable characterization of individuals' glycemic variation. However, there are concerns about their reliability for categorizing glycemic responses to foods that would limit their potential application in personalized nutrition recommendations.
Objectives:
We aimed to evaluate the concordance of 2 simultaneously worn CGM devices in measuring postprandial glycemic responses.
Methods:
Within ZOE PREDICT (Personalised Responses to Dietary Composition Trial) 1, 394 participants wore 2 CGM devices simultaneously [n = 360 participants with 2 Abbott Freestyle Libre Pro (FSL) devices; n = 34 participants with both FSL and Dexcom G6] for ≤14 d while consuming standardized (n = 4457) and ad libitum (n = 5738) meals. We examined the CV and correlation of the incremental area under the glucose curve at 2 h (glucoseiAUC0-2 h). Within-subject meal ranking was assessed using Kendall τ rank correlation. Concordance between paired devices in time in range according to the American Diabetes Association cutoffs (TIRADA) and glucose variability (glucose CV) was also investigated.
Results:
The CV of glucoseiAUC0-2 h for standardized meals was 3.7% (IQR: 1.7%-7.1%) for intrabrand device and 12.5% (IQR: 5.1%-24.8%) for interbrand device comparisons. Similar estimates were observed for ad libitum meals, with intrabrand and interbrand device CVs of glucoseiAUC0-2 h of 4.1% (IQR: 1.8%-7.1%) and 16.6% (IQR: 5.5%-30.7%), respectively. Kendall τ rank correlation showed glucoseiAUC0-2h-derived meal rankings were agreeable between paired CGM devices (intrabrand: 0.9; IQR: 0.8-0.9; interbrand: 0.7; IQR: 0.5-0.8). Paired CGMs also showed strong concordance for TIRADA with a intrabrand device CV of 4.8% (IQR: 1.9%-9.8%) and an interbrand device CV of 3.2% (IQR: 1.1%-6.2%).
Conclusions:
Our data demonstrate strong concordance of CGM devices in monitoring glycemic responses and suggest their potential use in personalized nutrition.This trial was registered at clinicaltrials.gov as NCT03479866.
Related Concept Videos
Diabetes: Symptoms, Diagnosis, and Complications
Diabetes Mellitus: Type 2 and Gestational
Overview of Carbohydrate Metabolism
Glucose transport into cells is facilitated by a family of transport proteins called GLUT (Glucose Transporters). GLUT4 is the primary glucose transporter for insulin-stimulated glucose...
Hypoglycemia and Glucagon
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...