AuraGain™ versus i-gel™ for bronchoscopic intubation under continuous oxygenation: A randomised controlled trial
Christine N Svendsen1, Charlotte V Rosenstock1,2, Gine L Glargaard1
1Department of Anaesthesiology and Intensive Care, Nordsjaellands Hospital, Hillerød, Denmark.
Introduction:
After failed mask ventilation and tracheal intubation, guidelines issued by the Difficult Airway Society recommend placing a second generation supraglottic airway device to secure oxygenation. Ultimately, a secure airway can be obtained by tracheal intubation through the supraglottic airway device using a bronchoscope. In this randomised trial, we compared the AuraGain™ with the i-gel™ as conduit for bronchoscopic intubation under continuous oxygenation performed by a group of anaesthesiologists with variable experience in a general population of patients.
Method:
We randomised one hundred patients who were equally allocated to flexible bronchoscopic intubation through the i-gel™ or the AuraGain™. In a random order, 25 anaesthesiologists each performed four intubations, two using the i-gel™ and two using the AuraGain™. Our primary outcome was 'total time for airway management'; i.e. total time from manually reaching the SAD to successful FBI confirmed at the end of the first inspiratory downstroke on the capnography curve.
Results:
In total, 87% (95% CI, 79%-92%) of the patients were successfully intubated through the allocated supraglottic airway device. There was no difference in total time for airway management between the i-gel™ and the AuraGain™ (199 vs. 227 s, p = .076). However, there was a difference in time for placement of the i-gel™, compared to the AuraGain™, (37 vs. 54 s, p < .001). There were nine failed intubations in the AuraGain™ group compared to four in the i-gel™ group (p = .147).
Conclusion:
We found no difference in total time for airway management between using the i-gel™ and using the AuraGain™.
More Related Videos
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
Related Concept Videos
Acute Respiratory Failure-IV
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques