Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Validation of an equine fitness tracker: ECG quality and arrhythmia detection
Fe Ter Woort1, Guillaume Dubois2, Grace Tansley3
1Equine Sports Medicine Practice, 83 avenue Beau Sejour, 1410, Waterloo, Belgium.
Background:
Cardiac arrhythmias in exercising horses are the focus of much interest, both in terms of what is considered normal and potential associations with poor performance and sudden cardiac death. One barrier to performing large-scale studies is the lack of an easily applicable device, to allow recording of large numbers of high-quality exercising electrocardiograms (ECGs). The EquimetreTM is a new wearable device which records a single lead ECG, amongst other parameters. Validation of such wearable devices is essential before further studies are undertaken.
Objectives:
To evaluate the quality of ECG using the EquimetreTM and compare arrhythmia detection during exercise with the reference TelevetTM system.
Study Design:
Prospective blinded clinical study.
Methods:
Simultaneous ECGs were recorded with both systems in 49 healthy horses during exercise. High intensity exercise (>40km/h) was performed by 29 racehorses, and lower intensity exercise for the remainder of the racehorses and show jumpers. Tracings were excluded if >10% artefact was present (duration of artefact relative to duration of exercise). For included ECGs, the duration of artefact was recorded and compared. ECGs were evaluated using Kubios premium software. Arrhythmia detection (yes/no) and arrhythmia classification (sinus arrhythmia, narrow complex of similar morphology to the sinus complexes or wide complex with a different morphology) were compared using Cohen's Kappa coefficient.
Results:
Nine TelevetTM ECGs and 3 EquimetreTM ECGs were excluded due to artefact >10%. TelevetTM ECGs included significantly more artefact during exercise than EquimetreTM ECGs (5% vs. 0.25% p<0.001). Arrhythmia analysis was performed on 38 horses' paired ECGs. The Kappa coefficient was excellent for arrhythmia detection (K=0.97) and arrhythmia classification (K=0.93).
Main Limitations:
Relatively low numbers of horses with arrhythmias (n=21) were included. The ECG recordings only provided one lead, making arrhythmia classification challenging in some cases.
Conclusions:
The Equimetre device provides a reliable ECG for arrhythmia detection during exercise. This system may be useful clinically and for future large-scale investigations into the occurrence and significance of exercising arrhythmias.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias