Preliminary study of automated oxygen titration at birth for preterm infants
Sanoj Km Ali1, Rohan V Jayakar2, Andrew P Marshall2
1Department of Paediatrics, Royal Hobart Hospital, Hobart, Tasmania, Australia.
Objective:
To study the feasibility of automated titration of oxygen therapy in the delivery room for preterm infants.
Design:
Prospective non-randomised study of oxygenation in sequential preterm cohorts in which FiO2 was adjusted manually or by an automated control algorithm during the first 10 min of life.
Setting:
Delivery rooms of a tertiary level hospital.
Participants:
Preterm infants <32 weeks gestation (n=20 per group).
Intervention:
Automated oxygen control using a purpose-built device, with SpO2 readings input to a proportional-integral-derivative algorithm, and FiO2 alterations actuated by a motorised blender. The algorithm was developed via in silico simulation using abstracted oxygenation data from the manual control group. For both groups, the SpO2 target was the 25th-75th centile of the Dawson nomogram.
Main Outcome Measures:
Proportion of time in the SpO2 target range (25th-75th centile, or above if in room air) and other SpO2 ranges; FiO2 adjustment frequency; oxygen exposure.
Results:
Time in the SpO2 target range was similar between groups (manual control: median 60% (IQR 48%-72%); automated control: 70 (60-84)%; p=0.31), whereas time with SpO2 >75th centile when receiving oxygen differed (manual: 17 (7.6-26)%; automated: 10 (4.4-13)%; p=0.048). Algorithm-directed FiO2 adjustments were frequent during automated control, but no manual adjustments were required in any infant once valid SpO2 values were available. Oxygen exposure was greater during automated control, but final FiO2 was equivalent.
Conclusion:
Automated oxygen titration using a purpose-built algorithm is feasible for delivery room management of preterm infants, and warrants further evaluation.
More Related Videos
05:45Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Oxygen Requirements and Growth Patterns