The Frontloading Fields Study: The Impact of False Positives and Seeding Point Errors on Visual Field Reliability
Jack Phu1,2, Michael Kalloniatis1,2
1Centre for Eye Health, University of New South Wales, Kensington, New South Wales, Australia.
Evaluating visual field (VF) reliability criteria, this study found that current false positive (FP) and seeding point error (SPE) thresholds may exclude valid data. Censoring low sensitivity results (≤19 dB) significantly improves intrasession VF correlations.
Area of Science:
- Ophthalmology
- Visual Science
- Medical Diagnostics
Background:
- Assessing visual field (VF) reliability is crucial for accurate glaucoma diagnosis and monitoring.
- Conventional criteria like false positives (FP) and seeding point errors (SPE) are used to determine VF result reliability.
- The impact of low sensitivity points on VF reliability assessments requires further investigation.
Purpose of the Study:
- To evaluate the impact of FP and SPE criteria on intrasession SITA-Faster VF result correlations.
- To assess the concurrent effect of low sensitivity points (≤19 dB) on VF reliability.
- To determine if censoring low sensitivity results improves VF correlation.
Main Methods:
- Analysis of 2320 intrasession SITA-Faster VF results from 1160 eyes.
- Categorization of results into 'both reliable' and 'reliable-unreliable' pairs.
- Evaluation of VF results against FP rates and SPE, with and without censoring sensitivity results ≤19 dB.
- Use of segmental linear regression to identify critical FP levels.
Main Results:
- A small increase in mean deviation (0.09 dB per 1% exceeding 12%) and sensitivity increase locations was observed with higher FP rates.
- SPEs primarily affected primary seeding points, with minimal impact on other VF indices.
- Censoring sensitivity results ≤19 dB significantly improved the correlation between reliable and unreliable VF results.
Conclusions:
- Current FP (>15%) and SPE criteria may lead to the erroneous exclusion of valid VF data.
- Censoring low sensitivity results (≤19 dB) enhances intrasession VF result correlations.
- Guidelines are provided for interpreting the impact of FP, SPE, and low sensitivity on VF analysis.