[Out-of-center blood pressure measurements in dialysis patients: Feasibility and comparison of methods]
Fatou Diawara1, Christine Fumeron1, Laurent Duflot1
1Hôpital privé du Vert Galant-Ramsay Santé, 38, rue du Dr Georges-Assant, 93290 Tremblay-en-France, France.
Insights
Out-of-center blood pressure monitoring is feasible in hemodialysis patients, with a 77.5% success rate. This supports the routine use of self-blood pressure measurement and ambulatory blood pressure monitoring in dialysis care.
Area of Science:
- Nephrology
- Hypertension Management
- Clinical Practice Guidelines
Background:
- European Renal Association-European Dialysis and Transplant Association (ERA-EDTA)/European Society of Hypertension (ESH) guidelines recommend out-of-center blood pressure measurements for dialysis patients.
- The practical feasibility of implementing these recommendations in routine clinical care remains largely unknown.
Purpose of the Study:
- To assess the "a priori" feasibility (proportion of patients to whom measurements can be proposed and who accept) and "a posteriori" feasibility (proportion of valid measurements obtained) of out-of-center blood pressure monitoring in hemodialysis patients.
- To evaluate the applicability of recommended blood pressure monitoring strategies in a real-world chronic hemodialysis setting.
Main Methods:
- A systematic program for out-of-center blood pressure measurements was conducted over six months in a chronic hemodialysis unit.
- Patients were offered either self-blood pressure measurement (SBPM) using an Omron M3 device or 44-hour ambulatory blood pressure monitoring (ABPM) using a Microlife WatchBP 03 device, based on eligibility criteria.
- SBPM involved morning and evening measurements over six non-dialysis days, while ABPM was performed between dialysis sessions with measurements every 15-30 minutes.
Main Results:
- Out of 129 hemodialysis patients, out-of-center measurements were not possible for 21 (16.3%).
- The "a priori" feasibility for SBPM was 62.8% (81/129), with 4 failures. The overall "a posteriori" feasibility for out-of-center measurements (SBPM and ABPM combined) was 77.5% (100/129).
- Ambulatory blood pressure monitoring was performed in 24 patients (23 ineligible for SBPM, 1 SBPM failure), with 19 valid measurements.
Conclusions:
- Out-of-center blood pressure measurements, including self-blood pressure measurement and ambulatory blood pressure monitoring, demonstrate good feasibility in the hemodialysis population.
- The study findings support the practical implementation of ERA-EDTA/ESH recommendations for blood pressure monitoring in routine dialysis care.
- These results indicate that out-of-center blood pressure monitoring can be successfully integrated into the management of hypertension in patients undergoing hemodialysis.
Abstract:
The European Renal Association-European Dialysis and Transplant Association (ERA-EDTA)/European Society of Hypertension (ESH) recommends out-of-center blood pressure measurements, self-blood pressure measurement or ambulatory blood pressure measurement in dialysis patients. However, the feasibility of out-of-center blood pressure measurements in routine care is not known. The objective of our study was to quantify it as "a priori" i.e. the percentage of hemodialysis to whom out-of-center blood pressure measurements can be proposed and who accept it, as "a posteriori", i.e. the percentage of out-of-center blood pressure measurements made and valid. A systematic out-of-center blood pressure measurements program was implemented from April to October 2019 in our chronic hemodialysis structures. It was proposed to each dialysis patient to carry out after education, an self-blood pressure measurement (Omron M3®), from 2 measurements, to 1 to 2minutes interval, mornings and evenings of 6days without dialysis (validity: 15 measures). Apart from arrhythmic patients, to all patients "not eligible" for self-blood pressure measurement (visually impaired, hemiplegic, neuropsychological disorders, language barrier), a 44-hour ambulatory blood pressure measurement (Microlife WatchBP 03®) was proposed separating 2 hemodialysis sessions; measures every 15minutes from 7 a.m. to 10 p.m. and 30minutes from 10 p.m. to 7 a.m. (validity: 40 measurements/day and 14/night). This is a study evaluating practices recommended for routine care in 18-year-old hemodialysis, having given their consent to the collection and analysis of the data. One hundred twenty nine patients were treated with chronic hemodialysis in our structures during the out-of-center blood pressure measurements campaign. Out-of-center blood pressure measurements could not be done in 21 patients (4 deceased, 2 transplanted and 4 absent before evaluation; 7 arrhythmics; 3 refusals and 1 multiple-disabled). Of these 108 patients (sex ratio 1.25; 69.3±13.5 years), 23 were ineligible for self-blood pressure measurement (visually impaired, neuro- and/or psychological disorders, language barrier). Due to 4 self-blood pressure measurement failures, the feasibility of the self-blood pressure measurement (n=81/129) is 62.8 % (CI95% 54.2-70.7). Of the 24 ambulatory blood pressure measurements performed (23 among those not eligible for self-blood pressure measurement and 1 failure of self-blood pressure measurement), 19 were valid. The "a posteriori" feasibility of out-of-center blood pressure measurements (n=100/129) is 77.5 % (CI95% 69.6-83.4). The feasibility of out-of-center blood pressure measurements in hemodialysis patients is good, making the application of the recommendations possible.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: