Related Experiment Video
Updated: Oct 3, 2025

Microsurgical Clip Obliteration of Middle Cerebral Aneurysm Using Intraoperative Flow Assessment
Published on: September 25, 2009
Pipeline Embolization Device and Pipeline Flex Versus Surpass Streamline Flow Diversion in Intracranial Aneurysms: A
Chaim M Feigen1, Juan Vivanco-Suarez2, Kainaat Javed1
1Montefiore Department of Neurological Surgery, Bronx, New York, USA.
Objective:
To compare safety and efficacy profiles in aneurysms treated with Pipeline Embolization Device or Pipeline Flex versus Surpass Streamline flow diverters (FDs).
Methods:
Patients who underwent flow diversion for aneurysm treatment at 2 centers were included. Covariates comprised patient demographics, comorbidities, and aneurysm characteristics. Metrics included number of devices, adjuvant device use, case duration, and radiation exposure. Outcomes included periprocedural complications and radiographic results at follow-up. Propensity score-matched pairs were generated using demographic and aneurysm characteristics to verify the outcomes in equally sized groups.
Results:
The majority of 141 flow diversion procedures performed on 126 patients were in the anterior circulation (96%) and unruptured (93%). Operators experienced more complications placing Surpass FDs compared with Pipelines (18.2% vs. 3.1%, P = 0.005) but used fewer Surpass devices per case (1 device in all Surpass cases and range for Pipeline cases 1-7; P < 0.001). Ballooning was more frequent for Surpass (29.5% vs. 2.1%, P < 0.001). There were no differences in mortality (2.1% vs. 0, P = 1.00), intracranial hemorrhage (3.1% vs. 0, P = 0.551), or stroke (4.2% vs. 6.8%, P = 0.680). Rates of aneurysm obliteration at follow-up were similar. Propensity-matched pairs had no differences in FD deployment complications or perioperative events, yet the significant differences remained for adjuvant balloon use and number of FDs deployed.
Conclusions:
While the devices demonstrated similar safety and efficacy profiles, deployment of the Surpass Streamline was more technically challenging than Pipeline Embolization Device or Pipeline Flex. Prospective cohort studies are needed to corroborate these findings.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
08:12Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023