A system for rapid analysis of the femoral blood velocity waveform at the bedside
Abstract:
The shape of the arterial blood velocity waveform varies with atherosclerotic disease and several methods of quantifying the shape in order to predict the severity of the disease have been described. These methods include pulsatility index, the Laplace transform method, and principal component analysis. This paper describes the development of a system which allows the operator to acquire, display, and store waveforms from each limb and then to quantify the waveforms at the bedside within a few minutes. The system includes a 10 MHz bi-directional Doppler unit, an instantaneous mean frequency processor, and an Apple II microcomputer fitted with an accelerator card. Both the Laplace transform parameters and the pulsatility index are computed and each result is printed in tabular form together with the averaged results of five waveforms from each limb. The printout is suitable for inclusion in the patient's folder. In initial clinical studies Laplace transform analysis exhibited a good correlation with aorto-iliac stenosis as assessed angiographically (R = 0.73 P less than 0.001 t test).