Optimal Design of an Ultrasound Transducer for Efficient Acousto-Optic Modulation of Terahertz Radiation
Pavel Alekseevich Nikitin1,2, Vasily Valerievich Gerasimov3,4
1Scientific and Technological Centre of Unique Instrumentation RAS, 117342 Moscow, Russia.
Abstract:
Acousto-optic (AO) interaction in the terahertz range was investigated with the use of monolithic ultrasound transducers of various widths. Sulfur hexafluoride (SF6) liquefied at a temperature of about 23 °C and a pressure of 25 bar was used as a medium for AO interaction. The angular and frequency bandwidths of effective AO interaction, as well as the diffraction efficiency per 1 W of the driving electric power, were determined. For the first time, a correct comparison of the AO diffraction efficiency in SF6 with the use of ultrasound transducers with different widths was carried out. In the experiments performed, the highest energy efficiency of the AO modulator was achieved with a transducer with a width of 12 mm.