Feasibility Study of Glucose Concentration Measurement of Aqueous Solution Using Time Domain Reflected Signals
Samira Saeedi1,2, Somayyeh Chammani1, Georg Fischer2
1Time-Domain Electromagnetics Laboratory, Faculty of Electrical Engineering, K.N. Toosi University of Technology, Tehran 1631714191, Iran.
Sensors (Basel, Switzerland)
|February 15, 2022
Summary
Time-domain wideband microwave spectroscopy (WBMS) offers a low-cost method for blood glucose sensing. This study demonstrates its effectiveness in measuring glucose concentration in aqueous solutions, showing a clear relationship between signal peaks and glucose levels.
Area of Science:
- Applied Physics
- Spectroscopy
- Biomedical Engineering
Background:
- Wideband microwave spectroscopy (WBMS) is utilized for material characterization.
- Traditional blood glucose sensing via microwave spectroscopy often employs resonant frequency-domain methods.
- Time-domain (TD) WBMS presents a cost-effective and accessible alternative for glucose sensing in aqueous solutions.
Purpose of the Study:
- To present early research on implementing a TD dielectric spectroscopy setup for glucose concentration measurement.
- To compare frequency-domain (FD) and TD methods for glucose sensing.
- To establish a relationship between TD reflected signals and glucose concentration.
Main Methods:
- Utilized a quasi-monostatic setup for measurements between 0.1 to 6 GHz.
- Calculated TD reflected signals from water with varying glucose content using inverse Laplace transform.
- Employed Vector Network Analyzer (VNA) with Vivaldi antennas for FD measurements and a maximum length sequence (m-sequence) transceiver for direct TD measurements.
Main Results:
- Measurement and numerical results showed good agreement, following the same trend.
- A monotonic relationship was observed between the peaks of TD signals and glucose concentration.
- The variation in reflected signal peak height was 0.00002 for FD and 0.0005 for direct TD measurements per 50 mg/dL glucose concentration.
Conclusions:
- The developed TD dielectric spectroscopy setup is effective for glucose concentration measurement.
- TD WBMS provides a viable and convenient method for blood glucose sensing.
- The study investigated glucose concentrations from 25 mg/dL to 400 mg/dL with a worst-case repeatability of 3.65%.