Evaluation of a crescent saw guide for tibial plateau-leveling osteotomy: An ex vivo study
Lindsay C Peterson1, Stanley E Kim1, Adam H Biedrzycki1
1Department of Small Animal Clinical Sciences, College of Veterinary Medicine, University of Florida, Gainesville, Florida, USA.
Objective:
To evaluate the effectiveness of a novel crescent-shaped tibial plateau-leveling osteotomy (TPLO) saw guide (crescent guide) to assist with saw control in novice participants.
Study Design:
Ex vivo study.
Sample Population:
Synthetic bones (n = 54) and medium sized dog pelvic limbs (n = 36).
Methods:
The 6 participants (interns and residents) without any prior experience performing a TPLO each performed 9 osteotomies on synthetic tibia models, and 6 osteotomies in cadaveric limbs of medium-sized dogs. Osteotomies made with the crescent guide were compared with those made with a standard jig and a radial saw guide with a jig. Osteotomy angulation, distance of eccentricity (DOE), and medial tibial cortical damage (synthetic bone models only) were measured from calibrated photographs. Participants rated their experiences with each technique.
Results:
There was no difference in the DOE, coronal or axial osteotomy angulation between the 3 alignment devices for synthetic bone models or cadavers. Average medial cortical damage with the crescent guide (3.8 ± 7.3 mm2 ) was lower than with the radial guide (35.7 ± 27 mm2 ) and standard jig (51.2 ± 63.2 mm2 ) guides (P = <.01). Five of 6 participants preferred the crescent guide over the standard jig and radial guide.
Conclusion:
There was no difference in accuracy of osteotomy positioning but using the crescent guide resulted in lower cortical damage and more favorable participant perceptions.
Clinical Relevance:
The crescent guide may improve control of the radial saw during TPLO in novice surgeons but does not appear to aid accurate osteotomy positioning.