Spatial temperature reconstructions in myocardial tissues undergoing radiofrequency ablations by performing
Martina Zaltieri1, Pietro Rossi2, Stefano Bianchi2
1Department of Engineering, Università Campus Bio-Medico Di Roma, Via Alvaro del Portillo, 00128, Rome, Italy.
Background:
Radiofrequency (RF) lesion creation is related to the heat propagation induced by RF application on tissues. Thermocouple embedded in the RF antenna are not able to predict deep tissue temperature at various level.
Objectives:
This study aims to investigate the influence of power delivered on radiofrequency catheter ablation (RFCA) effects by means of high resolved 2D temperature maps.
Methods:
Three trials of four ablations (12 applications) were executed on each specimen of healthy excised swine myocardium in different application points at four RF power values (30 W, 40 W, 50 W, and 60 W) for a fixed treatment time. All the data provided by the fiber Bragg gratings (FBGs) were analyzed. Temperature variations (ΔT) in time recorded in the 28 sites of measurements were reported. Also, temperature maps showing the ΔT spatial distribution reached within the tissue at the end of the RFCA were produced and displayed, together with the representation of the lethal isotherm. Moreover, the time of achievement of the lethal isotherm at different tissue depths (from 1 to 8 mm) was evaluated for the four power settings.
Results:
Temperature trends reported comparable profiles across the different power settings. ΔT values and ΔT rising times showed dependence on the sensors' proximity to the RF energy source and on the set RF power. Temperature maps confirmed that heat propagation occurs preferentially along the width of the tissue than in the depth. Also, for the adjusted treatment time, no power setting guarantees lesions thicker than 6 mm.
Conclusions:
ΔT maximal values and ΔT rising time strongly depends on the proximity of the tissues to RF energy source, as well as on the RF power setting. A plateau is reached in lesion size, regardless of the power setting. A first correlation between lesion size, power setting, and time to achieve lethal isotherms has been established.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...