Related Experiment Video
Updated: Oct 3, 2025

Efficient SARS-CoV-2 Quantitative Reverse Transcriptase PCR Saliva Diagnostic Strategy utilizing Open-Source Pipetting Robots
Published on: February 11, 2022
Investigation of discordant SARS-CoV-2 RT-PCR results using minimally processed saliva
Dawn White1, Jimmy Gu2, Catherine-Jean Steinberg3
1Biointerfaces Institute, McMaster University, Hamilton, ON, L8S 4L8, Canada.
Abstract:
Saliva is an attractive sample for coronavirus disease 2019 testing due its ease of collection and amenability to detect viral RNA with minimal processing. Using a direct-to-RT-PCR method with saliva self-collected from confirmed COVID-19 positive volunteers, we observed 32% false negative results. Confirmed negative and healthy volunteer samples spiked with 106 genome copies/mL of heat-inactivated severe acute respiratory syndrome coronavirus 2 showed false negative results of 10% and 13%, respectively. Additional sample heating or dilution of the false negative samples conferred only modest improvements. These results highlight the potential to significantly underdiagnose COVID-19 infections when testing directly from minimally processed heterogeneous saliva samples.