Phantom task-based image quality assessment of three generations of rapid kV-switching dual-energy CT systems on
Joël Greffier1, Anaïs Viry2,3, Yves Barbotteau4
1Department of Medical Imaging, CHU Nîmes, University of Montpellier, Nîmes Medical Imaging Group, Montpellier, France.
Purpose:
To compare the spectral performance of three rapid kV switching dual-energy CT (DECT) systems on virtual monoenergetic images (VMIs) at low-energy levels on abdominal imaging.
Methods:
A multi-energy phantom was scanned on three DECT systems equipped with three different gemstone spectral imaging (GSI) platforms: GSI (1st generation, GSI-1st), GSI-Pro (2nd generation, GSI-2nd ), and GSI-Xtream (3rd generation, GSI-3rd). Acquisitions on the phantom were performed with a CTDIvol close to 11mGy. For all platforms, raw data were reconstructed using filtered-back projection (FBP) and a hybrid iterative reconstruction algorithm (ASIR-V at 50%; AV50). A deep-learning image reconstruction (DLR) algorithm (TrueFidelity) was used only for the GSI-3rd. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated from 40 to 80 keV of VMIs. A detectability index (d') was computed to assess the detection of two contrast-enhanced lesions according to the keV level used.
Results:
For all GSI platforms, the noise magnitude decreased from 40 to 70 keV, and using AV50 compared to FBP. The average NPS spatial frequency (fav ) and spatial resolution (TTF50% ) were similar from 40 to 70 keV and decreased with AV50 compared to FBP. Compared to AV50, using DLR reduced the noise magnitude (-27% ± 3%) and improved fav values (10% ± 0%) and altering spatial resolution (2% ± 5%). For the two lesions, d' values peaked at 70 keV for GSI-1st and GSI-2nd platforms and at 40/50 keV for GSI-3rd, for all reconstruction algorithms. The highest d' values were found for the GSI-3rd with DLR.
Conclusion:
Differences in image quality were found between the GSI platforms for VMIs at low keV. The new DLR algorithm on the GSI-3rd platform reduced noise and improved spatial resolution and detectability without changing the noise texture for VMIs at low keV. The choice of the best energy level in VMIs depends on the platform and the reconstruction algorithm.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT