Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Fluoroless left atrial access for radiofrequency and cryoballoon ablations using a novel radiofrequency transseptal
Hany Demo1, Carla Aranda1, Mansour Razminia2
1Swedish Hospital, NorthShore University HealthSystem, Chicago, IL, USA.
Purpose:
Conventional catheter ablation for atrial fibrillation requires fluoroscopy, which has inherent risks of radiation exposure to patients and medical staff. Optimization of fluoroscopy parameters and use of three-dimensional electroanatomic mapping (EAM) and intracardiac echocardiography (ICE) have helped to reduce radiation exposure; however, despite growing evidence, there are still concerns about safety and added procedure time associated with fluoroless procedures, particularly in left-sided ablations, due to the potential risk of complications. Herein, we report our initial experience using a radiofrequency (RF) wire for completely fluoroless radiofrequency ablation (RFA) and cryoballoon ablation (CBA).
Methods:
A retrospective analysis was conducted on ablation procedures for various cardiac arrhythmias performed non-fluoroscopically at two centers using the VersaCross RF wire transseptal system under EAM and ICE guidance.
Results:
A total of 72 and 54 patients underwent RFA and CBA, respectively, successfully without any procedural complications. Transseptal access time for RFA was 14.5 ± 6.6 min from procedure start (including sheath and catheter placements ± right-sided ablation) or 2.8 ± 1.0 min from RF wire insertion into the femoral introducer. Transseptal access time for CBA was 19.2 ± 11.7 min from procedure start (including sheath and catheter placements ± right-sided ablation) or 3.5 ± 1.6 min from RF wire insertion into the femoral introducer. Average procedure time was 104.4 ± 38.0 min for RFA and 91.1 ± 22.1 min for CBA.
Conclusions:
A RF wire can be used to achieve completely fluoroless transseptal puncture safely and effectively while improving procedural efficiency in both RFA and CBA.