Related Experiment Video
Updated: Oct 2, 2025

Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Utilization of steerable sheath improves the efficiency of atrial fibrillation ablation guided by robotic magnetic
Qingzhi Luo1, Yun Xie1, Yangyang Bao1
1Department of Cardiovascular Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Background:
The objective of this study was to assess the impact of steerable sheaths compared with fixed-curve sheaths on the procedural outcomes of atrial fibrillation (AF) ablation guided by robotic magnetic navigation (RMN).
Methods And Results:
In this retrospective case-control study, 110 patients scheduled for AF catheter ablation were enrolled and divided into two groups. Fifty-five patients (paroxysmal, 70%) were treated with RMN-guided ablation utilizing a steerable sheath and another 55 patients (paroxysmal, 70%) were ablated with RMN using a fixed-curve sheath. Clinical characteristics were similar between the two groups. Compared with the fixed-curve sheath group, the steerable sheath group procedure time (111.9 ± 25.2 vs. 90.4 ± 20.7 min, p < .001) and radiofrequency (RF) time (35.9 ± 9.0 vs. 30.5 ± 7.4 min, p < .001) were significantly shortened. Additionally, the navigation index was significantly improved (0.41 ± 0.06 vs. 0.48 ± 0.08, p < .001) in the steerable sheath group. By employing a large catheter loop for targeting the right pulmonary veins (PVs), the steerable sheath group significantly reduced the RF delivery time (15.0 ± 3.0 vs. 12.0 ± 2.1 min, p < .001) during right-side PV isolation (PVI). However, total fluoroscopy time was similar between the two groups (5.6 ± 2.6 vs. 5.0 ± 2.0 min, p > .05). Acute PVI success rates were similar between the two groups. No major or minor complications occurred in either group.
Conclusion:
Appropriate utilization of steerable sheath technology can improve the efficiency of AF ablation guided by RMN, primarily by reducing the total procedure and RF delivery times of right-side PVI without compromising safety.