Related Experiment Video
Updated: Oct 2, 2025

ADSC-sheet Transplantation to Prevent Stricture after Extended Esophageal Endoscopic Submucosal Dissection
Published on: February 10, 2017
Lowering the setting value of the esophageal endoscopic submucosal dissection device enabled tissue damage control in
Yukiko Yamaguchi1, Masaya Uesato2, Shohei Yonemoto1
1Department of Frontier Surgery, Chiba University Graduate School of Medicine, Chiba, 260-8670, Japan.
Abstract:
One of the complications of esophageal endoscopic submucosal dissection (ESD) is postoperative stricture formation. Stenosis formation is associated with inflammation and fibrosis in the healing process. We hypothesized that the degree of thermal damage caused by the device is related to stricture formation. We aimed to reveal the relationship between thermal damage and setting value of the device. We energized a resected porcine esophagus using the ESD device (Flush Knife 1.5). We performed 10 energization points for 1 s, 3 s, and 5 s at four setting values of the device. We measured the amount of current flowing to the conducted points and the temperature and evaluated the effects of thermal damage pathologically. As results, the mean highest temperatures for 1 s were I (SWIFT Effect3 Wat20): 61.19 °C, II (SWIFT Effect3 Wat30): 77.28 °C, III (SWIFT Effect4 Wat20): 94.50 °C, and IV (SWIFT Effect4 Wat30): 94.29 °C. The mean heat denaturation areas were I: 0.84 mm2, II: 1.00 mm2, III: 1.91 mm2, and IV: 1.54 mm2. The mean highest temperature and mean heat denaturation area were significantly correlated (P < 0.001). In conclusion, Low-current ESD can suppress the actual temperature and thermal damage in the ESD wound.

