Concordance rate of radiologists and a commercialized deep-learning solution for chest X-ray: Real-world experience
Eun Young Kim1, Young Jae Kim2, Won-Jun Choi3
1Department of Radiology, Gil Medical Center, Gachon University College of Medicine, Incheon, South Korea.
Purpose:
Lunit INSIGHT CXR (Lunit) is a commercially available deep-learning algorithm-based decision support system for chest radiography (CXR). This retrospective study aimed to evaluate the concordance rate of radiologists and Lunit for thoracic abnormalities in a multicenter health screening cohort.
Methods And Materials:
We retrospectively evaluated the radiology reports and Lunit results for CXR at several health screening centers in August 2020. Lunit was adopted as a clinical decision support system (CDSS) in routine clinical practice. Subsequently, radiologists completed their reports after reviewing the Lunit results. The DLA result was provided as a color map with an abnormality score (%) for thoracic lesions when the score was greater than the predefined cutoff value of 15%. Concordance was achieved when (a) the radiology reports were consistent with the DLA results ("accept"), (b) the radiology reports were partially consistent with the DLA results ("edit") or had additional lesions compared with the DLA results ("add"). There was discordance when the DLA results were rejected in the radiology report. In addition, we compared the reading times before and after Lunit was introduced. Finally, we evaluated systemic usability scale questionnaire for radiologists and physicians who had experienced Lunit.
Results:
Among 3,113 participants (1,157 men; mean age, 49 years), thoracic abnormalities were found in 343 (11.0%) based on the CXR radiology reports and 621 (20.1%) based on the Lunit results. The concordance rate was 86.8% (accept: 85.3%, edit: 0.9%, and add: 0.6%), and the discordance rate was 13.2%. Except for 479 cases (7.5%) for whom reading time data were unavailable (n = 5) or unreliable (n = 474), the median reading time increased after the clinical integration of Lunit (median, 19s vs. 14s, P < 0.001).
Conclusion:
The real-world multicenter health screening cohort showed a high concordance of the chest X-ray report and the Lunit result under the clinical integration of the deep-learning solution. The reading time slight increased with the Lunit assistance.
More Related Videos
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...