Full-thickness endotracheal tube defect resulting in an anaesthetic circuit leak
1Royal (Dick) School of Veterinary Studies, University of Edinburgh, Roslin, Midlothian, UK.
Background:
Loss of endotracheal tube (ETT) integrity secondary to dental damage is reported in the human literature.
Objective:
To describe this problem in equine anaesthesia.
Study Design:
Case report.
Clinical Summary:
An 18-year-old Standardbred gelding presented out of hours with colic signs. Findings on clinical examination and pain refractory to analgesia meant that exploratory laparotomy was elected for. Prior to general anaesthesia (GA) leak testing of the anaesthetic machine was performed and the pilot balloon of the endotracheal tube (ETT) was inflated to confirm cuff integrity. Intermittent-positive pressure ventilation (IPPV) was initiated immediately following placement in dorsal recumbency and connection to the anaesthetic machine. During the inspiratory phase of IPPV, a loud gas leak was audible from the oropharynx and minimal thoracic excursion was observed, with repeated inflations of the ETT cuff unsuccessful at abolishing the leak. Due to suspicion of a defect within the silicone ETT itself, a support arm was used to abolish the curvature of the ETT, maintaining it in a straighter plane. This intervention abolished the leak allowing effective IPPV. After completion of GA, a close inspection of the ETT revealed a full-thickness laceration, thought to be a result of dental damage at an earlier date.
Main Limitations:
A single case is described.
Conclusions:
This report emphasises the importance of thorough inspection of the ETT prior to use to effectively secure the airway and enable IPPV provision in critical cases.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...