Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Influence of ambient temperature changes on intraoral scanning accuracy
Marta Revilla-León1, Aishwa Gohil2, Abdul B Barmak3
1Affiliate Assistant Professor Graduate Prosthodontics, Department of Restorative Dentistry, School of Density, University of Washington, Seattle, Wash; Director of Research and Digital Dentistry, Kois Center, Seattle, Wash; Adjunct Professor Graduate Prosthodontics, Department of Prosthodontics, School of Dental Medicine, Tufts University, Boston, Mass.
Statement Of Problem:
Different variables that decrease the accuracy of intraoral scanners (IOSs) have been identified. Ambient temperature changes can occur in the dental environment, but the impact of ambient temperature changes on intraoral scanning accuracy is unknown.
Purpose:
The purpose of this in vitro study was to assess the impact of ambient temperature changes on the accuracy (trueness and precision) of an IOS.
Material And Methods:
A complete arch maxillary dentate Type IV stone cast was obtained. Four 6-mm-diameter gauge balls were added to the maxillary cast to aid future evaluation measurements. The maxillary cast was digitized by using an industrial scanner (GOM Atos Q 3D 12M). The manufacturer's recommendations were followed in obtaining a reference scan. Then, the maxillary cast was digitized by using an IOS (TRIOS 4) according to the scanning protocol recommended by the manufacturer. Four groups were created depending on the ambient temperature change assessed: 24 °C or room temperature (24-D or control group), 19 °C or a 5-degree temperature drop (19-D group), 15 °C or a 9-degree temperature drop (15-D group), and 29 °C or a 5-degree temperature rise (29-D group). The Shapiro-Wilk and Kolmogorov-Smirnov tests revealed that the data were not normally distributed (P<.05). For trueness, the nonparametric Kruskal-Wallis followed by the Dwass-Steel-Critchlow-Fligner pairwise comparison tests were used. Precision analysis was obtained by using the Levene test based on the comparison of the standard deviations of the 4 groups with 95% Bonferroni confidence intervals for standard deviations (α=.05).
Results:
The Kruskal-Wallis test revealed significant differences in the trueness values among all 4 groups (P<.001). Furthermore, significant differences between the linear discrepancy medians between the control and 19-D groups (P<.001), control and 15-D groups (P=.002), control and 29-D groups (P<.001), 19-D and 29-D groups (P=.003), and 15-D and 29-D groups (P<.001) were found. The Levene test for the comparison of the variances among the 4 groups did not detect a significant difference (P>.999), indicating that precision wise the 4 groups were not significantly different from each other.
Conclusions:
Ambient temperature changes had a detrimental effect on the accuracy (trueness and precision) of the IOS tested. Ambient temperature changes significantly decreased the scanning accuracy of the IOS system tested. Increasing the ambient temperature has a greater influence on the intraoral scanning accuracy of the IOS selected when compared with decreasing the ambient temperature.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Factors Affecting Body Temperature
Factors may include: