Identifying the optimal monopolar electrocautery output power in pedicular internal thoracic artery harvesting: 20 or
Emin Can Ata1, Gözde Erkanli Şentürk2, Halil Ibrahim Saygi2
1Clinic of Cardiovascular Surgery, Haseki Education and Research Hospital, Istanbul, Turkey.
Background:
Monopolar electrocautery is an important tool for harvesting the pedicular internal thoracic artery (ITA) in cardiac surgery. The different power outputs of cautery may affect graft integrity and long-term patency. This study aimed to identify the optimal threshold of electrocautery power for ITA harvest.
Methods:
This prospective study included 30 patients who underwent elective coronary artery bypass surgery at the Medipol Mega University Hospital. The ITA was harvested by monopolar electrocautery after a median sternotomy. The output of cautery was adjusted at 20 W in group A and 40 W in group B. Three to 4 cm of a distal ITA sample from each patient was examined under a light microscope by two independent pathologists.
Results:
The ITA harvest time was longer in group A (21.2 ± 7.5 vs 10.3 ± 8.1 min, p < 0.001) than in group B. ITA free flow was similar in the two groups (43.6 ± 48.7 vs 51.7 ± 45.0 Ml/min, p = 0.762). Mild to moderate injury in the endothelial and sub-endothelial sample was more frequent in the low-cautery group (p = 0.0037).
Conclusions:
ITA endothelial integrity was found to be better preserved with 40W electrocautery. Moreover, 20W of monopolar electrocautery may not be safe in pedicular ITA harvesting.