Related Experiment Video
Updated: Oct 2, 2025

Rapid Quantification of Mitogen-induced Blastogenesis in T Lymphocytes for Identifying Immunomodulatory Drugs
Published on: December 27, 2016
Validation of automated fluorescent-based technology for measuring total nucleated cell viability of hematopoietic
Kanan A Patel1, Nita Patel1, Grace H Kim1
1Department of Stem Cell Transplantation and Cellular Therapy, John Theurer Cancer Center, Hackensack University Medical Center, Hackensack, New Jersey, USA.
Background:
A reliable rapid method for measuring total nucleated cell (TNC) viability is essential for cell-based products manufacturing. The trypan blue (TB) exclusion method, commonly used to measure TNC viability of hematopoietic progenitor cell (HPC) products, is a subjective assay, typically uses a microscope, and includes a limited number of cells. The NucleoCounter NC-200 is an automated fluorescent-based cell counter that uses pre-calibrated cartridges with acridine orange and DAPI dyes to measure cell count and viability. This study describes the validation of the NC-200 for testing HPC's viability.
Methods:
Samples from 189 fresh and 60 cryopreserved HPC products were included. Fresh products were tested for viability after collection by both TB and NC-200. 7-aminoactinomycin D (7AAD) CD45+ cell viability results were obtained from a flow cytometry test. Cryopreserved products thawed specimens were tested for viability by both TB and NC-200. The NC-200 viability results were compared with the other methods. Acceptability criteria were defined as ≤10% difference between the NC-200 method and the other methods for at least 95% of the samples.
Results:
Fresh products' mean viability difference between NC-200 and TB or 7AAD CD45+ method was 4.9% (95%CI 4.6-5.4) and 2.8% (95%CI 2.2-3.4), respectively. Thawed products' mean viability difference between NC-200 and TB was 3.0% (95%CI 0.4-5.6).
Conclusion:
The NC-200 automated fluorescent-based method can be used effectively to determine HPC's viability for both fresh and cryopreserved products. It can help eliminate human bias and provide consistent data and operational ease.
More Related Videos
07:57Measurement of Mitochondrial Mass and Membrane Potential in Hematopoietic Stem Cells and T-cells by Flow Cytometry
Published on: December 26, 2019
10:20Simultaneous Assessment of Kinship, Division Number, and Phenotype via Flow Cytometry for Hematopoietic Stem and Progenitor Cells
Published on: March 24, 2023