Is the image quality of conventional chest radiography obtained from a two-layer flat panel detector affected by the
Shinya Takarabe1, Taku Kuramoto2, Yusuke Shibayama3
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi-ku, Fukuoka 812-8582, Japan; Department of Oral and Maxillofacial Radiology, Faculty of Dental Science, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka 812-8582, Japan.
Purpose:
Recently developed and commercialized dual-layer flat panel detectors (DL-FPDs) with two indirect scintillators are capable of acquiring dual-energy X-ray images. However, in clinical practice, they are utilized to perform conventional radiography using diagnostic X-rays with a wide energy spectrum. The two layers of the DL-FPD may affect the obtained image quality, even when only using one layer for conventional image acquisition, and these effects are yet to be substantiated. Therefore, in this study, we quantitatively evaluated the image quality of a conventional chest radiography using DL-FPD and visually verified the characteristics of the chest anthropomorphic phantom images.
Methods:
The physical characteristics of the system were evaluated using the pre-sampled modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE), for beam quality RQA 7 and RQA 9. In addition, the subjective visibility of the anthropomorphic chest phantom and simple objects images were compared with those of a conventional single-layer flat-panel detector (SL-FPD).
Results:
No significant differences were found in the MTF between the SL-FPD and DL-FPD images. In addition, a higher DQE was observed at some exposure doses and in the high spatial frequency regions wherein NNPSs were lower for DL-FPD than for SL-FPD. Furthermore, no significant differences were found in the subjective visibility of the chest phantoms in each system.
Conclusions:
We concluded that the image quality of the conventional radiography acquired with DL-FPD is comparable to or better than that of the SL-FPD.
More Related Videos
13:44Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Ultrasonography
During an ultrasonography procedure, a handheld device called...