Survivorship of a Metal-on-Metal Total Hip Implant With Modular Titanium Adapter
Related Experiment Video
Updated: Oct 2, 2025

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Derek J Semaan1, Heath Rutledge-Jukes2, Keith R Berend1
1Joint Implant Surgeons Inc, New Albany, OH; White Fence Surgical Suites, New Albany, OH; Mount Carmel Health System, New Albany, OH.
Background:
Use of metal-on-metal (MoM) articulations in total hip arthroplasty (THA) has sharply declined due to high failure rates from metal-related complications. Although certain MoM designs have demonstrated only 46% survival, not all MoM designs have performed the same. The purpose of this study is to evaluate mid-term to long-term survival of a specific MoM implant with a modular titanium taper adapter.
Methods:
A retrospective review was performed on all patients who underwent primary THA at our center with the M2a-Magnum system (Zimmer Biomet, Warsaw, IN). Of 829 patients (956 hips) identified, 754 patients (869 hips) met inclusion criteria of signed research consent, minimum 2-year follow-up, and/or any revision surgery.
Results:
Mean follow-up was 11.0 years (range 2-16; ±3.5). Mean cup angle of inclination was 42.8° (range 24°-70°, ±6.3°), with 88.0% reconstructed within the 40° ± 10° safe zone. There were 64 revisions (7.36%): 7 (0.81%) septic and 57 (6.56%) aseptic. Of those, 32 (3.68%) were adverse reactions to metal debris. Kaplan-Meier survival free of revision for all causes was 88.6% at 16 years (95% confidence interval 86.8-90.4). Univariate analysis of risk factors for all-cause, aseptic, and adverse reaction to metal debris revision found no relationship with female gender, age ≥65 years, body mass index >30 kg/m2, higher activity level, or inclination angle outlier.
Conclusion:
The results of this study demonstrate a more favorable mid-term to long-term survivorship with this specific MoM implant compared to other designs. Although our institution no longer performs MoM THA, further investigation into differences in MoM implant designs is warranted.