Oxygenator impact on peramivir in extra-corporeal membrane oxygenation circuits
Jeffrey J Cies1,2,3, Wayne S Moore1, Daniel Marino2
1The Center for Pediatric Pharmacotherapy LLC, Pottstown, PA, USA.
Introduction:
This study aims to determine the oxygenator impact on alterations of peramivir (PRV) in a contemporary neonatal/pediatric (1/4-inch) and adolescent/adult (3/8-inch) extra-corporeal membrane oxygenation (ECMO) circuit including the Quadrox-i® oxygenator.
Methods:
1/4-inch and 3/8-inch, simulated closed-loop ECMO circuits were prepared with a Quadrox-i pediatric and Quadrox-i adult oxygenator and blood primed. Additionally, 1/4-inch and 3/8-inch circuits were also prepared without an oxygenator in series. A one-time dose of PRV was administered into the circuits and serial pre- and post-oxygenator concentrations were obtained at 5-min and 1-, 2-, 3-, 4-, 5-, 6-, 8-, 12-, and 24-h time points. PRV was also maintained in a glass vial, and samples were taken from the vial at the same time periods for control purposes to assess for spontaneous drug degradation.
Results:
For the 1/4-in. circuit with an oxygenator, there was < 15% PRV loss, and for the 1/4-in. circuit without an oxygenator, there was < 3% PRV loss during the study period. For the 3/8-in. circuits with an oxygenator, there was < 15% PRV loss, and for the 3/8-in. circuits without an oxygenator, there was < 3% PRV loss during the study period.
Conclusion:
There was no significant PRV loss over the 24-h study period in either the 1/4-in. or 3/8-in circuit, regardless of the presence of the oxygenator. The concentrations obtained pre- and post-oxygenator appeared to approximate each other, suggesting there may be no drug loss via the oxygenator. This preliminary data suggests PRV dosing may not need to be adjusted for concern of drug loss via the oxygenator. Additional single and multiple dose studies are needed to validate these findings.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Transport in the Blood
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...