Pneumococcal IgG Antibody Responses to 23vPPV in Healthy Controls Using an Automated ELISA
Laine M Hosking1,2, Kirsten P Perrett3,4,5, Christine Czajko6
1Immunology Laboratory, Laboratory Services, Royal Children's Hospital, Melbourne, Australia. laine.hosking@rch.org.au.
Journal of Clinical Immunology
|March 1, 2022
Summary
This study verified interpretation criteria for an automated pneumococcal ELISA, finding it effective for measuring antibody responses to the 23-valent pneumococcal polysaccharide vaccine (23vPPV) in adults and children.
Area of Science:
- Immunology
- Vaccinology
- Clinical Diagnostics
Background:
- Measuring antibody responses to the 23-valent pneumococcal polysaccharide vaccine (23vPPV) is crucial for assessing vaccine efficacy.
- Current gold-standard methods like the World Health Organization ELISA are complex and labor-intensive.
- Interpretation criteria for an adequate response to 23vPPV require further validation.
Purpose of the Study:
- To verify the interpretation criteria for an in-house, automated multi-serotype ELISA used by the diagnostic Immunology Laboratory at The Royal Children's Hospital (RCH).
- To assess the immunogenicity of 23vPPV in both pneumococcal conjugate vaccine (PCV)-naïve adults and PCV-primed children.
Main Methods:
- Healthy adults (n=40) and children (n=22) received 23vPPV immunization.
- Serum samples were collected pre- and post-immunization (28-42 days).
- Samples were analyzed using an automated Tecan Freedom Evo 200 ELISA, with specific criteria defining an adequate antibody response.
Main Results:
- The RCH interpretation criteria were verified, with 97.5% of adults and 100% of children achieving an adequate response.
- A notable finding was that serotypes included in PCV were less immunogenic in PCV-naïve adults, with 30% not meeting response criteria based on these serotypes alone.
- The automated ELISA demonstrated high concordance with established criteria.
Conclusions:
- The study successfully validated the RCH interpretation criteria for their automated multi-serotype pneumococcal ELISA.
- The findings support the use of this automated method for reliable assessment of 23vPPV immunogenicity.
- Consideration of PCV serotype-specific responses is important for interpreting vaccine immunogenicity in different populations.
Related Concept Videos
Pneumonia III: Complications and Assessment
509
Pneumonia poses the potential for numerous complications that warrant consideration. These complications include the following:
509
Enzyme-Linked Immunosorbent Assay
14.7K
In 1971, Peter Perlman and Eva Engvall developed an Enzyme-linked immunosorbent assay (ELISA or EIA). ELISA differs from western blot in that the assays are conducted in microtiter plates or in vivo rather than on an absorbent membrane.
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
14.7K