Related Experiment Video
Updated: Oct 1, 2025

Preparing an Isotopically Pure 229Th Ion Beam for Studies of 229mTh
Published on: May 3, 2019
Optimal design of high-voltage DC power supply of 1.2 MV/45 mA applied to boron neutron capture therapy system
Sun-Hong Min1, Chawon Park1, Sukhwal Ma1
1Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul 01812, South Korea.
Abstract:
To build a proton beam accelerator that can be applied to a boron neutron capture therapy system based on an electrostatic accelerator, a high-voltage direct-current (DC) power supply system equivalent to the generation of neutrons should be provided. The symmetrical Cockcroft-Walton voltage multiplier method is suitable for stable acceleration of the proton beam in the tandem electrostatic accelerator in this system. Before the second step-up with the Cockcroft-Walton circuit, the design of the inverter is prioritized by preponderantly considering the first voltage and resonance frequency. Moreover, the optimized stacking number is determined with consideration of the ripple voltage, voltage drop, average output voltage, and fundamental harmonics, and a design is performed to set related parameter values to be stable in the flat-top region of the voltage. A high-voltage DC power supply system of 1.2 MV/45 mA is needed for a stable terminal energy of 2.4 MeV/20 mA. Such a design can be optimized by securing reliable data using a simulation tool on the basis of theoretical calculations. This will become a formidable touchstone in manufacturing technology based on acquiring practical know-how for setting up a tandem electrostatic accelerator-based boron neutron capture therapy system in the future.