Automatic contouring using deformable image registration for tandem-ring or tandem-ovoid brachytherapy
Yagiz Yedekci1, Melis Gültekin1, Sezin Yuce Sarı1
1Department of Radiation Oncology, Faculty of Medicine, Hacettepe University, Ankara, Turkey.
Deformable image registration (DIR)-based automatic contouring is effective for cervical image-guided brachytherapy (IGBT) using tandem-ring or tandem-ovoid applicators. This method significantly reduces contouring time, proving safe and efficient for clinical use.
Area of Science:
- Radiation Oncology
- Medical Imaging
- Gynecologic Oncology
Background:
- Image-guided brachytherapy (IGBT) is crucial for cervical cancer treatment.
- Accurate contouring of targets and organs at risk (OARs) is essential for optimal dose delivery.
- Manual contouring is time-consuming and subject to inter-observer variability.
Purpose of the Study:
- To evaluate the efficacy of deformable image registration (DIR)-based automatic contouring in 3D computed tomography (CT)-based IGBT for cervical cancer.
- To compare DIR-based automatic contouring with manual contouring for tandem-ring (T-R) and tandem-ovoid (T-O) applicators.
- To assess the dosimetric impact of DIR-based contouring on treatment plans.
Main Methods:
- Retrospective analysis of CT images from 28 cervical cancer patients undergoing IGBT with T-R or T-O insertion.
- Hybrid DIR performed between the first fraction CT and subsequent CTs, using the first CT as the reference.
- Evaluation of contour similarity using Dice Similarity Coefficient (DSC) and comparison of mean volumes; dosimetric analysis of dose-volume parameters.
Main Results:
- DIR-based automatic contours generally showed larger volumes than manual contours, with statistical significance for the small bowel.
- Lower DSC scores were observed for the small bowel and sigmoid.
- DIR-based contouring was less accurate for the rectum with T-O applicators compared to T-R.
- Volume differences did not significantly impact dose-volume parameters, and treatment plans covered DIR contours.
- DIR significantly reduced contouring time (2.0 ±0.1 min) compared to manual contouring (14.0 ±0.4 min) (p < 0.001).
Conclusions:
- DIR-based automatic contouring is a successful and safe method for T-R and T-O applications in cervical IGBT.
- The technique significantly decreases contouring time, enhancing efficiency in clinical practice.
- Automatic contouring using DIR is a time-saving and viable approach for image-guided brachytherapy.