Related Experiment Video
Updated: Oct 1, 2025

Endoscopic Vacuum Therapy for the Treatment of Anastomotic Leakage after Total Gastrectomy with Esophagojejunostomy
Published on: August 22, 2025
An unusual cause of failure in Zenith Alpha Abdominal endograft
Raffaella N Berchiolli1, Michele Marconi2,3, Irene Bargellini4
1Vascular Surgery Unit, Department of Translational Research and New Technologies in Medicine and Surgery, University of Pisa, Pisa, Italy.
Background:
Graft disruption is an unusual complication of the endovascular abdominal aortic aneurysm repair (EVAR).
Case Presentation:
A 71-year-old man underwent standard EVAR with Zenith Alpha Abdominal endograft. Follow-up examinations revealed an initial significant sac shrinkage. At 24 months, duplex ultrasound (DUS) scan and computed tomography showed increase of the sac diameter associated with complete disconnection of the suprarenal stent-graft from the main body without evidence of endoleak. A standard relining with a thoracic endograft was performed between the suprarenal stent and the main body of the previous graft. At 6 months DUS revealed sac shrinkage.
Conclusions:
This report demonstrates an uncommon cause of endograft failure with suprarenal stent disconnection from main body and highlights the need for continuous follow-up in patients undergoing EVAR.