Related Experiment Video
Updated: Oct 1, 2025

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Impact of lung block shape on cardiac dose for total body irradiation
Mallory C Glenn1, Kent Wallner1, Samuel M H Luk1
1Department of Radiation Oncology, University of Washington, Seattle, WA, United States.
Abstract:
Evaluating cardiac dose during total body irradiation (TBI) is of increasing interest. A three-dimensional beam model for TBI was commissioned and lung shielding was simulated in a treatment planning system with the cardiac silhouette partially blocked and unblocked. When blocked, the median heart dose decreased by 6% (IQR = 6%) and the median cardiac V12Gy decreased by 27% (IQR = 17%). The median left anterior descending artery dose decreased 20% (IQR = 12%) for blocked cases. Because using partial heart shielding may result in considerable changes in dose to cardiac structures, TBI protocols should explicitly consider lung block design parameters and their potential effects.

