A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Novel ventricular tachyarrhythmia detection enhancement detects undertreated life-threatening arrhythmias
Bruce L Wilkoff1, Laurence D Sterns2, Michael S Katcher3
1Sydell and Arnold Miller Family Heart, Vascular, and Thoracic Institute - Robert and Suzanne Tomsich, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio.
Background:
Ventricular tachyarrhythmias (VTA) with low and varying signal amplitudes and morphologies may not be successfully identified utilizing traditional implantable cardioverter-defibrillator algorithms.
Objective:
Develop and validate a novel algorithm (VF Therapy Assurance, VFTA) to improve detection and timely delivery of high-voltage therapy (HVT) for these arrhythmias.
Methods:
Arrhythmia detection was simulated on recorded VTA electrograms (EGMs) utilizing Abbott's Merlin.net database. EGMs where an HVT occurred only when VFTA was enabled, or where VFTA provided an HVT >30 seconds earlier than without VFTA, were readjudicated with physician review. As VFTA never prevents detection or therapy, EGMs where VFTA did not activate or alter HVT were not adjudicated.
Results:
Among 564,353 recorded VTA EGMs from 20,000 devices, VFTA altered HVT in 105 EGMs from 67 devices. Physician adjudication determined that 81.9% (86/105) of these EGMs were true undertreated VTA episodes and would have received appropriate HVT with VFTA enabled. Furthermore, 65% of the episodes (56/86) were ventricular fibrillation, were polymorphic, did not self-terminate during the recording window, or were not amenable antitachycardia pacing. Of those, 87.5% (49/56) would not have elicited HVT without VFTA. Overall, VFTA provided new or earlier appropriate HVT in 0.27% (53/20,000) of devices with an increase in inappropriate HVT in 0.07% (14/20,000) devices.
Conclusion:
The VFTA algorithm successfully identifies VTA missed by traditional detection algorithms, owing to undersensed ventricular signals resulting in the rate falling below the programmed detection rate. The use of VFTA increases the likelihood of delivering life-saving HVT.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...