Suppression of electromagnetic interference in electroretinography from a patient with an implanted left ventricular
Rebecca M Schur1,2, Brendan L Eck3, Alex Yuan4,5
1Cole Eye Institute, Cleveland Clinic, Cleveland, OH, USA. schurr@ccf.org.
Purpose:
A left ventricular assist device (LVAD) is an implantable cardiac pump that uses a magnetically-levitating rotor to pump blood into circulation for patients with congestive heart failure. The continuous high-frequency motion of the pump can cause significant interference in electroretinography (ERG) recordings. We evaluate filtering methods to improve ERG quality in the presence of LVAD interference.
Methods:
A patient with an implanted LVAD was referred to our clinic for ERG testing on suspicion of a retinal dystrophy. Full-field ERG (ffERG) and pattern ERG (pERG) were performed according to ISCEV standards. Recordings were acquired once in full-bandwidth mode and again in low-bandwidth mode. Digital low-pass and band-stop filtering were performed to mitigate ERG interference. Post-processing was also evaluated in a control subject with no implanted device.
Results:
High-frequency interference was present in all ERG recordings and corresponded to the speed settings of the pump. When applied in post-processing, both low-pass and band-stop filters suppressed the interference and presented readable ERGs without affecting peak times or amplitudes. By contrast, when recording in low-bandwidth mode, the filter drop-off was not steep enough to completely remove the interference and peak delays were introduced that could not be readily corrected.
Conclusions:
LVAD interference in ERG waveforms can be successfully removed using simple digital filters. If post hoc data processing capabilities are unavailable, a large amount of interference can be removed by narrowing the acquisition bandwidth and averaging additional repeats of each stimulus response.